8545-M838-O1 Rev D 825 Installation & Technical
SETUP AND CALIBRATION, CONT.
Total - Calibration – Filtering
Zero Track:
With the Total – Calibration – Filtering screen displayed the current setting for the Zero
Track: parameter will be shown. If the setting displayed is acceptable, proceed to the
next parameter.
Otherwise, press the SPACE key to step through and select the zero tracking range and
then press the Down Arrow to advance to the next parameter or press the ENTER key
to save the new setting and proceed to the next screen. Available selections are
Off, 0.5 div., 1 div., 2 div., 3 div., 4 div., 5 div., 6 div., 7 div., 8 div., and 9 div.
Motion Range:
With the Total – Calibration – Filtering screen displayed the current setting for the
Motion Range: parameter will be shown. If the setting displayed is acceptable,
proceed to the next parameter.
Otherwise, use the numeric keys to enter the motion range (the number of divisions of
change permitted before indicating unstable) and then press the Down Arrow to
advance to the next parameter or press the ENTER key to save the new setting and
proceed to the next screen. Allowable values are 0 to 9. Note that entering zero (0)
disables the motion range.
Filtering:
With the Total – Calibration – Filtering screen displayed the current setting for the
Filtering: parameter will be shown. If the setting displayed is acceptable, press the
ENTER key to proceed to the next screen or the Navigation Keys Left Arrow to return
to the previous screen.
Otherwise, press the SPACE key to step through and select the digital filtering and then
press the ENTER key to save the new setting and proceed to the next screen. Allowable
selections are Off, Minimal, Moderate, and Custom.
Minimal Filtering (Sample Rate = 2)
Moderate Filtering (Sample Rate = 1)
Custom Filtering (Additional Parameters Displayed)
NOTE: When Custom Filtering (Custom) has been selected, the following
parameters, TYPE:, Filter Break: and Filter Value: will be displayed.