8545-M838-O1 Rev D 825 Installation & Technical
SETUP AND CALIBRATION, CONT.
Intervals:
Three
If Three is selected, four (4) additional parameters will be displayed on the screen.
Press the Down Arrow to advance to the;
Low Interval Capacity:
This is the weight capacity for the low range interval. If the setting displayed is
acceptable, press the Down Arrow to advance to Middle Interval Value:.
Otherwise, use the numeric keys to enter the weight capacity for the low range interval
and then press the Down Arrow to advance to the next parameter. Note that the value
entered cannot exceed 7 digits and must be less than the overall scale capacity.
Middle Interval Value:.
This is the middle range interval value. If the setting displayed is acceptable, press the
Down Arrow to advance to the Middle Interval Capacity:.
Otherwise, use the numeric keys to enter the middle interval weight increment value and
then press the Down Arrow to advance to the next parameter. Note that the value
entered must be bigger than the low interval value.
Middle Interval Capacity:
This is the weight capacity for the middle range interval. If the setting displayed is
acceptable, press the Down Arrow to advance to the High Interval Value:.
Otherwise, use the numeric keys to enter the weight capacity for the middle range
interval and then press the Down Arrow to advance to the next parameter. Note that
the value entered cannot exceed 7 digits and must be less than the overall capacity and
greater than the low capacity.
High Interval Value:
This is the middle range interval value. If the setting displayed is acceptable, press the
ENTER key to save the setting and proceed to the Totalizer screen.
Otherwise, use the numeric keys to enter the high interval weight increment value (the
value entered must be bigger than the middle interval value) and then press the ENTER
key to save the new setting and proceed to the Totalizer screen.
In either case, after pressing the ENTER key, the display will briefly show:
Setting Parameters and then Setting Multi-Interval
and then change to the Totalizer screen.
NOTE: The error message Invalid three interval settings will be
displayed after pressing the ENTER key, if incorrect values have been entered for
Low Interval Capacity:, Middle Interval Value:, Middle
Interval Capacity: and High Interval Value:. To change the settings,
press the /ESC key to return to the Intervals: screen.