8545-M838-O1 Rev D 825 Installation & Technical
73
SETUP AND CALIBRATION, CONT.
Calibration from the Setup Menu
1. From the Startup screen, press the 3 key or use the Navigation Keys to select (highlight)
3. Legal metrology information / setup option and then press the ENTER key.
2. From the Legal metrology information / Setup screen, press the 3 key or use the Navigation
Keys to select (highlight) 3. Setup menu and then press the ENTER key.
3. Enter the Login and Password if the Login screen is displayed.
4. With the Setup Menu screen displayed, press the 3 key or use the Navigation Keys to
select (highlight) 3. Calibration and then press the ENTER key.
5. With the Setup Scale Number screen displayed, verify the correct scale number is shown
and then press the ENTER key. Otherwise, use the numeric keys to enter the scale
number and then press the ENTER key.
6. The display will change to show the Scale 1 Calibration Options screen, press
the 2 key or use the Navigation Keys to select (highlight) 2. Calibration and then
press the ENTER key.
7. The display will change to show the Scale 1 Calibration Weight 1 screen.
8. If this is the zero weight (no load), press the 0 key and then press the ENTER key.
Otherwise, place the weights on the scale and then using the numeric keys, enter the
current weight on the scale and then press the ENTER key.
9. The display will show the percentage of completion and a dashed line proceeding from left
to right for this step. When this step has been completed, the display will change to show
the Scale 1 Calibration Weight 2 screen.
10. If this is the zero weight (no load), press the 0 key and then press the ENTER key.
Otherwise, place the weights on the scale and then using the numeric keys, enter the
current weight on the scale and then press the ENTER key.
11. The display will show the percentage of completion and a dashed line proceeding from left
to right for this step. When this step has been completed, the display will change to change
to show the Scale 1 Calibration Weight 3 screen.
12. If this is the zero weight (no load), press the 0 key and then press the ENTER key.
Otherwise, place the weights on the scale and then using the numeric keys, enter the
current weight on the scale and then press the ENTER key.
13. The display will show the percentage of completion and a dashed line proceeding from left
to right for this step. When this step has been completed, the display will change to show
the Scale 1 Fine Span screen. NOTE: If Span Adjustment is required, proceed to
the FINE SPAN ADJUSTMENT.