8545-M838-O1 Rev D 825 Installation & Technical
SETUP AND CALIBRATION, CONT.
Scale 1 – Fine Span
If Fine Span adjustment is desired, with the Scale 1 – Fine Span screen displayed,
place a calibrated test weight on the scale.
Use the Navigation Keys to select (highlight) the Fine Span adjustment option and then
press the ENTER key to perform the function or;
Press the U key (U. Up) to increase displayed weight.
Press the D key (D. Down) to decrease the displayed weight,
Press the Z key (Z. Zero) to zero the scale.
Press the Q key (Q. Quit) to save the current settings and Quit.
The display will return to the Scale 1 – Calibration Options screen.
If 2nd Order Linearization was enabled and the Num Cal Pts parameter is set for 2
or more, then a COEFF option will be added to the Fine Span screen. This option is used to
make adjustments to the displayed weight after linearization testing has shown an error in a
specific range of weight.
Press the C key (C. Coeff) or use the Navigation Keys to select (highlight) the C. Coeff
option and then press the ENTER key to proceed to the Coefficient screen.
Linear Coefficient:
The Linear Coefficient is applied to the load cell signal to calculate weight. If the current
setting displayed for this coefficient is acceptable, press the Down Arrow to advance to
the 2nd Order Coeff: parameter.
Otherwise, use the numeric keys to enter a new Linear Coefficient and then press the
Down Arrow to advance to the next parameter.
2nd Order Coeff:
This coefficient is applied to the square of the load cell signal to calculate weight. If the
current 2nd Order Coeff setting displayed is acceptable, press the Down Arrow to
advance to the Tweak low weight: parameter.
Otherwise, use the numeric keys to enter a new 2nd Order Coeff and then press the
Down Arrow to advance to the next parameter.
WARNING! Changing the Linear Coefficient or the 2nd Order Coeff
will affect the indicator's ability to display the proper weight. Please
consult the factory before making any changes.