Chapter 5 – Troubleshooting
Section 5.3 Error Messages
5 - 13
Drawer Message This occurs if: Service Light
Open the door and
drawer and complete
entry of treatment data.
Touch ‘OK’ to continue or
‘Cancel’ to quit.
The drawer is closed before the
treatment data has been entered.
Open the drawer and enter the
treatment data.
On after
‘Cancel’ is
selected
Open the door and
drawer. Make sure there
are no sets in the
illuminator. Close the
drawer and door when
finished. Touch ‘OK’ to
continue or ‘Cancel’ to
quit.
Processing set(s) are in the
illuminator and should not be.
Remove the processing set(s) and
continue. If this message
continues to occur, contact your
authorized service representative.
On after
‘Cancel’ is
selected
Close the drawer. Touch
‘OK’ to continue or
‘Cancel’ to quit.
The drawer is open and it should
be closed. Close the drawer and
continue. If this message
continues to occur, contact your
authorized service representative.
Off
Open the door, and
push the drawer into the
latched position. Close
the door. Touch ‘OK to
continue or ‘Cancel’ to
quit.
The drawer is unlatched and it
should be latched. Close the
drawer and continue. If this
message continues to occur,
contact your authorized service
representative.
On after
‘Cancel’ is
selected
MAN-EN 00601, v3.0

Table of Contents