Chapter 5 – Troubleshooting
Section 5.3 – Error Messages
5 - 15
Front Door Message This occurs if: Service Light
Close the door and
drawer. Touch ‘OK’ to
continue or ‘Cancel’ to
quit.
The door and/or drawer are
open and should be closed.
Close the drawer and door to
continue. If this message
continues to occur, contact your
authorized service representative.
On after
‘Cancel’ is
selected
The door is unlocked.
Open and close the
door. Touch ‘OK’ to
continue or ‘Cancel’ to
quit.
The door is unlocked and should
be locked. Open and close the
door to continue. If this message
continues to occur, contact your
authorized service representative.
On after
‘Cancel’ is
selected
Open the door and
drawer and complete
entry of treatment data.
Touch ‘OK’ to continue or
‘Cancel’ to quit.
The drawer is closed before the
treatment data has been
entered. Open the drawer and
enter the treatment data.
On after
‘Cancel’ is
selected
Open the door and
drawer. Make sure there
are no sets in the
illuminator. Close the
drawer and door when
finished. Touch ‘OK’ to
continue or ‘Cancel’ to
quit.
Processing set(s) are in the
illuminator and should not be.
Remove the processing set(s)
and continue. If this message
continues to occur, contact your
authorized service representative.
On after
‘Cancel’ is
selected
The door did not lock.
Open and close the
door. Touch ‘OK’ to
continue or ‘Cancel’ to
quit.
The door is unlocked and should
be locked. Open and close the
door to continue. If this message
continues to occur, contact your
authorized service representative.
On after
‘Cancel’ is
selected
The door is open. Open
and close the door.
Touch ‘OK’ to continue or
‘Cancel’ to quit.
The door is open and should be
closed. Close the drawer and
door to continue. If this message
continues to occur, contact your
authorized service representative
for assistance.
On after
‘Cancel’ is
selected
The door is locked and
should not be. Touch
‘Cancel’ to quit. Contact
service for assistance.
The door is locked and should be
unlocked. Try gently pushing on
the door to release the lock. If
that does not work and there are
processing sets in the illuminator,
turn the illuminator off to release
the lock. If this message
continues to occur, contact your
authorized service representative
for assistance.
On