Chapter 5 – Troubleshooting
Section 5.3 – Error Messages
5 - 16
Integrity
Check Issues
(Self-tests)
These are the self-tests that the illuminator runs each time the power is
turned on or every 24 hours. These tests ensure that the device is working
properly.
Message This occurs if: Service Light
An Integrity Check Issue
has occurred (C0002).
Contact service for
assistance.
A problem is detected during
the self-tests. Contact your
authorized service
representative for assistance.
Off
An Integrity Check Issue
has occurred (C0003).
Contact service for
assistance.
A problem is detected during
the self-tests. Contact your
authorized service
representative for assistance.
On
An Integrity Check Issue
has occurred (C0004).
Contact service for
assistance.
A problem is detected during
the self-tests. Ensure the side
access panel is closed. Turn the
illuminator off, wait until the
screen darkens, then turn on to
restart the illuminator. If this
message continues to occur,
contact your authorized service
representative for assistance.
On
An Integrity Check Issue
has occurred (C0005).
Contact service for
assistance.
A problem is detected during
the self-tests. Turn off the
illuminator and open the door.
Open and close the drawer.
Close the door and turn on the
illuminator. If this message
continues to occur, contact
your authorized service
representative for assistance.
On
An Integrity Check Issue
has occurred (C0006).
Contact service for
assistance.
A problem is detected during
the self-tests. Turn off the
illuminator. Open and close the
door, then turn on the
illuminator. If this message
continues to occur, contact
your authorized service
representative for assistance.
On
MAN-EN 00601, v3.0