Chapter 5 – Troubleshooting
Section 5.3 – Error Messages
5 - 17
Integrity
Check Issues
(Self-tests)
Message This occurs if: Service Light
An Integrity Check Issue
has occurred (C0007).
Contact service for
assistance.
A problem is detected during
the self-tests. Turn off the
illuminator. Ensure no
processing sets are in the
drawer. Close the drawer and
door, then turn on the
illuminator. If this message
continues to occur, contact
your authorized service
representative for assistance.
On
An Integrity Check Issue
has occurred (C0008).
Contact service for
assistance.
A problem is detected during
the self-tests. Ensure the air filter
is clean and present. See
Section 6.3 for cleaning
instructions. If this message
continues to occur, contact
your authorized service
representative for assistance.
On
An Integrity Check Issue
has occurred (C0009).
Contact service for
assistance.
A problem is detected during
the self-tests. Turn off the
illuminator. Open and close the
door, then turn on the
illuminator. If this message
continues to occur, contact
your authorized service
representative for assistance.
On
An Integrity Check Issue
has occurred (C0010).
Contact service for
assistance.
A problem is detected during
the self-tests. The agitator is
stopped or is moving slowly.
Contact your authorized
service representative for
assistance.
On
An Integrity Check Issue
has occurred (C0011).
Contact service for
assistance.
A problem is detected during
the self-tests. Contact your
authorized service
representative for assistance.
On
An Integrity Check Issue
has occurred (C0012).
Contact service for
assistance.
A problem is detected during
the self-tests. Contact your
authorized service
representative for assistance.
On