STP-iX457/iXU457 Series Turbomolecular Pump
MT-86E-001-F
Page 143
7
SAFETY FUNCTIONS
Referred
Section
6
6
The magnetic bearing output stops and the rotor is descended on the touch down bearing.
The driver output stops and the rotor continue rotating by inertia. It may take several hours to stop.
To stop the STP pump quickly, close the vacuum valve at the outlet port flange and introduce gas from the purge port into the STP pump.
Table 43 - Error List (continued)
Countermeasures
rvice office.
Check whether or not a
power failure has occurred.
Connect the power cable
correctly.
Replace the power cable.
Set the rated voltage ±
10 %.
Probable causes
Failure of the STP pump.
Failure of the control unit.
isconnection of the motor
sensor.
control unit.
Power failure.
Incorrect connection of the
power cable.
Failure of the power cable.
Failure of the power supply
vlotage.
Occurrence
condition
c bearing
motor
Pump operation
*1
ecelerate and
*2
ecelerate and
(Error message)
nordinate
, 4