Configureprocessmeasurement
Table5-12Standardconcentrationcurvesandassociatedmeasurementunits
continued
CurvenameDescriptionDensityunitTemperature
unit
Concentration
unit
HFCS90Curverepresentsahydrometerscale
forHFCS90(highfructosecorn
syrup)solutionsthatindicatesthe
percentbymassofHFCSinsolution.
g/cm3°C%
5.10.3Derivedvariablesandcalculatedprocessvariables
Whenyouconguretheconcentrationmeasurementapplication,yourchoiceofDerivedVariable
determinestheprocessvariablesthatwillbecalculatedbytheapplication.
Table5-13liststheoptionsforDerivedVariable,andthesetofprocessvariablesthatarecalculated
foreachoption.
Table5-13Derivedvariablesandcalculatedprocessvariables
Calculatedprocessvariables
DerivedVariableDescription
Densityat
reference
tempera-
ture
Standard
volume
flowrate
Specific
gravity
Concen-
tration
Netmass
flowrate
Net
volume
flowrate
Densityatreference
temperature
Mass/unit
volume,
correctedtoa
givenreference
temperature
üü
SpecicgravityTheratioof
thedensity
ofaprocess
uidatagiven
temperatureto
thedensityof
wateratagiven
temperature.
Thetwogiven
temperature
conditionsdo
notneedtobe
thesame.
üüü
Massconcentration
derivedfrom
referencedensity
Thepercent
massofsolute
orofmaterialin
suspensionin
thetotalsolution,
derivedfrom
referencedensity
üüüü
96
MicroMotion9739MVDTransmitters

Table of Contents