Integratethemeterwiththecontrolsystem
Beforeassigningactionstoanenhancedeventordiscreteinput,checkthestatusofthe
eventortheremoteinputdevice.IfitisON,allassignedactionswillbeperformedwhenthe
newcongurationisimplemented.Ifthisisnotacceptable,waituntilanappropriatetimeto
assignactionstotheeventordiscreteinput.
Procedure
SetDiscreteInputActionasdesired.
ThedefaultsettingisNone.
OptionsforDiscreteInputAction
OptionsforDiscreteInputActionareshowninTable7-11.
Table7-11OptionsforDiscreteInputActionorEnhancedEventAction
Label
ActionDisplayProLinkIIFieldCommunicator
None(default)NONE
NoneNone
StartsensorzeroSTARTZEROStartSensorZeroPerformautozero
Start/stopalltotalizersSTARTSTOPStart/StopAllT otalizationStart/stoptotals
Resetmasstotal
RESETMASS
ResetMassT otalResetmasstotal
Resetvolumetotal
RESETVOL
ResetVolumeT otalResetvolumetotal
Resetgasstandardvolume
total
RESETGSVTResetGasStdVolume
Total
Resetgasstandardvolume
total
Resetalltotals
RESETALL
ResetAllTotalsResettotals
Resettemperature-
correctedvolumetotal
TCVOLResetAPIRefVolT otal
Resetcorrectedvolume
total
ResetCMreference
volumetotal
RESETSTDVResetCMRefVolT otalN/A
ResetCMnetmasstotalRESETNETMResetCMNetMassTotalN/A
ResetCMnetvolumetotalRESETNETVResetCMNetVolT otalN/A
IncrementCMmatrixINCrCURVEIncrementCurrentCM
Curve
N/A
Beforeassigningactionstoanenhancedeventordiscreteinput,checkthestatusofthe
eventortheremoteinputdevice.IfitisON,allassignedactionswillbeperformedwhenthe
newcongurationisimplemented.Ifthisisnotacceptable,waituntilanappropriatetimeto
assignactionstotheeventordiscreteinput.
7.4.2ConfigureDiscreteInputPolarity
Display
OFF-LINEMAINT→OFF-LINECONFG→IO→DI→DIPOLAR
ProLinkII
ProLink→Conguration→DiscreteInput→DI1Polarity
FieldCommunicatorCongure→ManualSetup→Inputs/Outputs→DiscreteInput→Polarity
ConfigurationandUseManual
135