Transmitteroperation
Procedure
Toviewthecurrentstatusofallpossiblealarms,selectProLink→Status.T oviewstatusindicatorsin
analarmcategory,clickononeofthreetabsintheStatuswindow:Critical,Informational,orOperational.
TheStatuswindowdisplaysthecurrentstatusofallpossiblealarms,includingIgnorealarmsonthe
threepanels.Atabisredifoneormorestatusindicatorsinthatcategoryis“on.”
AgreenLEDindicatesanInactivealarm.
AredstatusindicatorindicatesanActivealarm.
ToviewandacknowledgeactiveandinactiveFaultandInformationalarms,selectProLink→AlarmLog.
Toacknowledgeanalarm,checktheACKcheckbox.
ThetransmitterautomaticallyltersoutIgnorealarms.Entriesinthealarmlogaredividedintotwo
categories:HighpriorityandLowpriority.Theseprioritiescorrespondtothedefaultseveritylevels
forthetypeofalarm.Withineachcategory:
AredLEDindicatesanActivealarm.
AgreenLEDindicatesanInactivealarmthathasnotbeenacknowledged.
8.4.3ViewalarmsusingtheFieldCommunicator
Procedure
PressServiceTools→AlertstoviewtheactiveFaultandInformationalalarms.Thetransmitter
automaticallyltersoutIgnorealarms.
PressServiceTools→Alerts→RefreshAlertstorefreshtheviewoftheactivealarms.
PressServiceTools→Alerts,andselecttheindividualalarmtoviewandacknowledgethealarm.
8.4.4InteractionoftransmitterfaultactionsandAlarmSeveritylevel
Whenanalarmconditionoccurs,thetransmitterrespondsbyreportingastatusalarm.Eachstatus
alarmhasanalarm.Alarmsareclassiedintothreeseveritylevels:Fault,Information,andIgnore.Severity
levelcontrolshowthetransmitterrespondstothealarmcondition.YoucanchangetheAlarmSeverity
forsomealarms.
Thetransmittermaintainstwostatusagsforeachalarm:
TherststatusagindicatesActiveorInactive.
ThesecondstatusagindicatesAcknowledgedorUnacknowledged.
Whenthetransmitterdetectsanalarmcondition,thefollowingoccurs:
Analarmispostedforthecorrespondingalarm:
FirststatusagissettoActive.
SecondstatusagissettoUnacknowledged.
Thetransmittercheckstheseveritylevelforthespecicalarm:
IfSeverityisFault,outputsgototheirconguredFaultAction(aftertheconguredfaulttimeout
hasexpired).
IftheSeverityisInformationorIgnore,outputsarenotaffected.Theycontinuetoreportprocess
data.