EasyManua.ls Logo

Emerson Micro Motion 9739 MVD

Emerson Micro Motion 9739 MVD
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
Table10-3Flowproblemsandrecommendedactions
continued
ProblemPossiblecausesRecommendedactions
Erraticnon-zeroow
rateunderno-ow
conditions
Leakingvalveorseal
Slugow
Pluggedowtube
Incorrectsensororientation
Wiringproblem
Vibrationinpipelineatratecloseto
sensortubefrequency
Dampingvaluetoolow
Mountingstressonsensor
Sensorcross-talk
Verifythatthesensororientationis
appropriateforyourapplication(referto
thesensorinstallationmanual).
Checkthedrivegainandthepickoff
voltage.SeeSection10.25andSection
10.26.
Forinstallationwith9-wirecabling,
verifythatthe9-wirecableiscorrectly
grounded.
Checkthewiringbetweenthesensorand
transmitter.SeeSection10.10.
Forsensorswithajunctionbox,check
formoistureinthejunctionbox.
Purgetheowtubes.
Checkforopenorleakingvalvesor
seals.
Checkforsourcesofvibration.
Verifydampingconguration.
Verifythatthemeasurementunitsare
conguredcorrectlyforyourapplication.
Checkforslugow.SeeSection10.24.
Checkforradiofrequencyinterference.
SeeSection10.12.
Iftwosensorswithsimilarfrequencyare
tooneareachother,separatethem.
ContactMicroMotion.
ConfigurationandUseManual
185

Table of Contents

Related product manuals