QuickstartwithProLinkII
Figure3-4ProLinkIIconnectiontomultidropnetwork
B
C
D
A
AHARTinterface(toPC)
B250–600Ωresistance
CDevicesonthenetwork
DMasterdevice
Notes
Thisgureshowsaserialportconnection.AUSBportconnectionisalsoavailable.
4.StartProLinkII.
5.ChooseConnection→ConnecttoDevice.
6.SetProtocoltoHARTBell202.
Tip
HART/Bell202connectionsusestandardconnectionparameters.Youdonotneedtocongure
themhere.
7.IfyouareusingaUSBportconnection,enableConverterTogglesRTS.
8.SetAddress/TagtotheHARTpollingaddressconguredinthetransmitter.
32MicroMotion9739MVDTransmitters

Table of Contents