EasyManua.ls Logo

Emerson Micro Motion 9739 MVD - Page 38

Emerson Micro Motion 9739 MVD
254 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
QuickstartwithProLinkII
Figure3-4ProLinkIIconnectiontomultidropnetwork
B
C
D
A
AHARTinterface(toPC)
B250–600Ωresistance
CDevicesonthenetwork
DMasterdevice
Notes
Thisgureshowsaserialportconnection.AUSBportconnectionisalsoavailable.
4.StartProLinkII.
5.ChooseConnection→ConnecttoDevice.
6.SetProtocoltoHARTBell202.
Tip
HART/Bell202connectionsusestandardconnectionparameters.Youdonotneedtocongure
themhere.
7.IfyouareusingaUSBportconnection,enableConverterTogglesRTS.
8.SetAddress/TagtotheHARTpollingaddressconguredinthetransmitter.
32MicroMotion9739MVDTransmitters

Table of Contents

Related product manuals