Cerabar M PMC51, PMP51, PMP55
Endress+Hauser 63
Universal process adapter
68.5 (2.7)
1
4.2
5
6
4.1
5
6
2
3
ø43.5 (1.71)
= ø26 (1)
d
M
A0023744
1 Top section: AISI 316L (1.4404)
2 Bottom section: AISI 316L (1.4435)
3 Slotted-nut: AISI 304 (1.4301) or AISI 304L (1.4307)
4.1 Form seal: silicone or EPDM
4.2 Profile seal: EPDM and stainless steel
5 Measuring cell seal
6 Ceramic sensor
Engineering unit mm (in)
The roughness of the surface in contact with the medium is R
a
< 0.76 µm (30 µin).
Silicone form seal: FDA 21CFR177.2600/USP Class VI, order number: 52023572
EPDM form seal: FDA, USP Class VI; 5 pcs, order number: 71100719
EPDM and stainless steel profile seal: FDA, USP Class VI; 1 pc, order number: 71431380
Designation Nominal pressure
bar (psi)
Weight Process connection approval Option
1)
[kg (lb)]
Universal process adapter
Form seal made of silicone (4.1)
10 (145) 0.74 (1.63) ASME-BPE
For more, see Product Configurator
UPJ
Universal process adapter
Form seal made of
EPDM (4.1) or
EPDM and stainless steel (4.2)
ASME-BPE
For more, see Product Configurator
UNJ
1) Product Configurator, order code for "Process connection"
Material of the form seal (exchangeable
seal)
Material of the measuring cell seal on
ceramic sensor (seal not exchangeable)
Approval of measuring cell seal Option
1)
Silicone EPDM FDA
2)
3A Class II, USP Class VI. DVGW, KTW,
W270, WRAS, ACS, NSF61
K
EPDM or
EPDM and stainless steel
EPDM FDA
2)
J
1) Product Configurator, order code for "Seal"
2) Food-safe FDA 21 CFR 177.2600

Table of Contents