EasyManua.ls Logo

Endress+Hauser HART Proline t-mass I 300 - Diagnostic message

Endress+Hauser HART Proline t-mass I 300
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline t-mass I 300 HART Diagnostics and troubleshooting
Endress+Hauser 139
11.3 Diagnostic information on local display
11.3.1 Diagnostic message
Faults detected by the self-monitoring system of the measuring device are displayed as a
diagnostic message in alternation with the operational display.
Operational display in alarm condition Diagnostic message
1
4
3
X X X X X X XX X X X X X X X XX X
2
S
S
XX
20.50
5
X
i
S801
Menu
Supply voltage
A0029426-EN
1
2
3
4
5
Status signal
Diagnostic behavior
Diagnostic behavior with diagnostic code
Short text
Operating elements
If two or more diagnostic events are pending simultaneously, only the message of the
diagnostic event with the highest priority is shown.
Other diagnostic events that have occurred can be displayed in the Diagnostics menu:
Via parameter →  148
Via submenus →  149
Status signals
The status signals provide information on the state and reliability of the device by
categorizing the cause of the diagnostic information (diagnostic event).
The status signals are categorized according to VDI/VDE 2650 and NAMUR
Recommendation NE 107: F = Failure, C = Function Check, S = Out of Specification, M
= Maintenance Required
Symbol Meaning
Failure
A device error has occurred. The measured value is no longer valid.
Function check
The device is in service mode (e.g. during a simulation).
Out of specification
The device is operated:
Outside its technical specification limits (e.g. outside the process temperature range)
Outside of the configuration carried out by the user (e.g. maximum flow in parameter
20 mA value)
Maintenance required
Maintenance is required. The measured value remains valid.

Table of Contents

Related product manuals