Proline t-mass I 300 HART Commissioning
Endress+Hauser 83
Parameter overview with brief description
Parameter Prerequisite Description Selection / User entry
Reference conditions – Select reference conditions for
calculation of the corrected volume
flow.
• 1013.25 mbara, 0 °C
• 1013.25 mbara, 15 °C
• 1013.25 mbara, 20 °C
• 1013.25 mbara, 25 °C
• 1000 mbara, 0 °C
• 1000 mbara, 15 °C
• 1000 mbara, 20 °C
• 1000 mbara, 25 °C
• 14.696 psia, 59 °F
• 14.696 psia, 60 °F
• User-defined
Reference pressure The Others option is selected in the
Reference conditions parameter.
Select reference conditions for the
corrected volume flow.
0 to 250 bar a
Reference temperature – Enter reference temperature for
calculating the reference density.
–200 to 450 °C
FAD conditions The Air or compressed air option is
selected in the Measurement
application parameter parameter.
Select reference conditions for the
calculation of the FAD density (FAD =
free air delivery).
• 1000 mbara, 20 °C
• 14.504 psia, 68 °F
• User-defined
FAD pressure • The Air or compressed air option is
selected in the Measurement
application parameter parameter.
• The User-defined option is selected
in the FAD conditions parameter
parameter.
Enter reference pressure for the
calulation of the FAD density (FAD =
free air delivery).
0 to 250 bar a
FAD temperature • The Air or compressed air option is
selected in the Measurement
application parameter parameter.
• The User-defined option is selected
in the FAD conditions parameter
parameter.
Enter reference temperature for the
calulation of the FAD density (FAD =
free air delivery).
–200 to 450 °C
Reference combustion temperature The Energy option is selected in the
Measurement application parameter
parameter.
Enter reference combustion
temperature to calculate the natural
gas energy value.
–200 to 450 °C
9.4.4 Sensor adjustment
Parameters pertaining to the pipe shape of the insertion version can be configured in the
Sensor adjustment submenu.
The number of submenus and parameters can vary depending on the device version.
Certain submenus and parameters in these submenus are not described in the
Operation Instructions. Instead a description is provided in the Special Documentation
for the device (→ "Supplementary documentation" section).
‣
Sensor adjustment
Installation direction
→ 84
Installation factor
→ 84
Pipe shape
→ 84
Pipe inner diameter
→ 84