Commissioning Proline t-mass I 300 HART
102 Endress+Hauser
9.4.13 Configuring the low flow cut off
The Low flow cut off wizard systematically guides the user through all the parameters
that must be set to configure low flow cut off.
Navigation
"Setup" menu → Low flow cut off
‣
Low flow cut off
Assign process variable
→ 102
On value low flow cutoff
→ 102
Off value low flow cutoff
→ 102
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry
Factory setting
Assign process variable – Select process variable for low
flow cut off.
• Off
• Mass flow
• Volume flow
• Corrected volume
flow
• FAD volume flow
*
–
On value low flow cutoff A process variable is selected
in the Assign process variable
parameter (→ 102).
Enter on value for low flow cut
off.
Positive floating-
point number
Depends on country
and nominal
diameter
Off value low flow cutoff A process variable is selected
in the Assign process variable
parameter (→ 102).
Enter off value for low flow cut
off.
0 to 100.0 % –
* Visibility depends on order options or device settings