Operation options Proline t-mass I 300 HART
64 Endress+Hauser
Working area
Depending on the selected function and the related submenus, various actions can be
performed in this area:
Configuring parameters
Reading measured values
Calling up help text
Starting an upload/download
7.4.6 Disabling the Web server
The Web server of the measuring device can be switched on and off as required using the
Web server functionality parameter.
Navigation
"Expert" menu → Communication → Web server
Parameter overview with brief description
Parameter Description Selection
Web server functionality Switch the Web server on and off. Off
HTML Off
On
Function scope of the "Web server functionality" parameter
Option Description
Off The web server is completely disabled.
Port 80 is locked.
On The complete functionality of the web server is available.
JavaScript is used.
The password is transferred in an encrypted state.
Any change to the password is also transferred in an encrypted state.
Enabling the Web server
If the Web server is disabled it can only be re-enabled with the Web server functionality
parameter via the following operating options:
Via local display
Via Bedientool "FieldCare"
Via "DeviceCare" operating tool
7.4.7 Logging out
Before logging out, perform a data backup via the Data management function
(upload configuration from device) if necessary.
1. Select the Logout entry in the function row.
The home page with the Login box appears.
2. Close the Web browser.
3. If no longer needed:
Reset modified properties of the Internet protocol (TCP/IP) →  60.