Commissioning Proline t-mass I 300 HART
82 Endress+Hauser
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
Mol% Ne – Enter amount of substance for
the gas mixture.
Ne = neon
0 to 100 % –
Mol% NH3 – Enter amount of substance for
the gas mixture.
NH
3
= ammonia
0 to 100 % –
Mol% O2 – Enter amount of substance for
the gas mixture.
O
2
= oxygen
0 to 100 % –
Mol% O3 As a mixture, only possible
with O2:
• O3: 0 to 35 %
• O2: 65 to 100 %
O3 as single gas:
100 %
Enter amount of substance for
the gas mixture.
0 to 100 % –
Mol% Xe – Enter amount of substance for
the gas mixture.
Xe = xenon
0 to 100 % –
Special gas name Special gas option application
package is available.
Shows the description of the
gas ordered by the customer,
e.g. gas name or gas
composition.
- -
* Visibility depends on order options or device settings
9.4.3 Configuring reference conditions
Reference properties can be configured in the Reference conditions submenu.
Navigation
"Setup" menu → Reference conditions
‣
Reference conditions
Reference conditions
→ 83
Reference pressure
→ 83
Reference temperature
→ 83
FAD conditions
→ 83
FAD pressure
→ 83
FAD temperature
→ 83
Reference combustion temperature
Reference combustion temperature
→ 83