Diagnostics and troubleshooting Proline t-mass I 300 HART
152 Endress+Hauser
Info number Info name
I1626 Write protection deactivated
I1627 Web server: login successful
I1628 Display: login successful
I1629 CDI: login successful
I1631 Web server access changed
I1632 Display: login failed
I1633 CDI: login failed
I1634 Reset to factory settings
I1635 Reset to delivery settings
I1639 Max. switch cycles number reached
I1649 Hardware write protection activated
I1650 Hardware write protection deactivated
I1712 New flash file received
I1725 Sensor electronic module (ISEM) changed
I1726 Configuration backup failed
11.11 Resetting the measuring device
Using theDevice reset parameter (→ 113) it is possible to reset the entire device
configuration or some of the configuration to a defined state.
11.11.1 Function scope of the "Device reset" parameter
Options Description
Cancel No action is executed and the user exits the parameter.
To delivery settings Every parameter for which a customer-specific default setting was ordered is reset
to this customer-specific value. All other parameters are reset to the factory
setting.
Restart device The restart resets every parameter whose data are in the volatile memory (RAM)
to the factory setting (e.g. measured value data). The device configuration remains
unchanged.
Restore S-DAT backup Restore the data that are saved on the S-DAT. The data record is restored from the
electronics memory to the S-DAT.
This option is displayed only in an alarm condition.
11.12 Device information
The Device information submenu contains all parameters that display different
information for device identification.
Navigation
"Diagnostics" menu → Device information
‣
Device information
Device tag
→ 153