Installation Proline t-mass I 300 HART
28 Endress+Hauser
5.2 Mounting the measuring device
5.2.1 Required tool
For sensor
Sensor's compression fitting: appropriate mounting tool.
5.2.2 Preparing the measuring device
1. Remove all remaining transport packaging.
2. Remove any protective covers or protective caps present from the sensor.
3. Remove stick-on label on the electronics compartment cover.
5.2.3 Mounting the measuring device
L
WARNING
Danger due to improper process sealing!
Ensure that the seals are clean and undamaged.
Ensure that the correct sealing material is used (e.g. Teflon tape for NPT compression
fitting).
Secure the seals correctly.
Install the measuring device or turn the transmitter housing so that the cable entries do
not point upwards.
A0029263
1. Weld in the mounting boss in accordance with the requirements.
2. Loosen the union nut (1) and push down the compression fitting (2).
3 mm6
1.
3 mm
2.
3.
A0041022