Device Functions Proline Promass 83 MODBUS RS485 5 Block USER INTERFACE
Endress+Hauser 45
ASSIGN
(continued)
2420 Advanced selection: with the optional SW package CONCENTRATION
61 = TARGET MASS FLOW
62 = % TARGET MASS FLOW
63 = TARGET VOLUME FLOW
64 = % TARGET VOLUME FLOW
65 = CORRECTED TARGET VOLUME FLOW
66 = CARRIER MASS FLOW
67 = % CARRIER MASS FLOW
68 = CARRIER VOLUME FLOW
69 = % CARRIER VOLUME FLOW
70 = CORRECTED CARRIER VOLUME FLOW
71 = % BLACK LIQUOR
72 = ° BAUME
74 = ° API
75 = ° PLATO
76 = ° BALLING
77 = ° BRIX
78 = OTHER (_ _ _ _ flexible concentration)
Advanced selection: with the optional SW package ADVANCED DIAGNOSIS
79 = MASS FLOW DEVIATION
80 = DENSITY DEVIATION
81 = REFERENCE DENSITY DEVIATION
82 = TEMPERATURE DEVIATION
83 = TUBE DAMPING DEVIATION
84 = ELECTRODYNAMIC SENSOR DEVIATION
96 = OPERATING FREQUENCY FLUCTUATION DEVIATION
97 = TUBE DAMPING FLUCTUATION DEVIATION
! Note!
Multiplex mode is suspended as soon as a fault / notice message is generated. The
message in question appears on the display.
Once the fault is eliminated, the measuring device resumes operation in Multiplex
mode and the error message is no longer displayed on the local display.
100% VALUE
MODBUS register:
Data type:
Access:
2421
2534
Float
read/write
! Note!
Function is not available unless one of the following was selected in the ASSIGN
function (2420):
•MASS FLOW IN %
•VOLUME FLOW IN %
• CORRECTED VOLUME FLOW IN %
• MASS FLOW BARGRAPH IN %
• VOLUME FLOW BARGRAPH IN %
• CORRECTED VOLUME FLOW BARGRAPH IN %
For specifying the flow value to be shown on the display as the 100% value.
User input: 5-digit floating-point number
Factory setting: 10 kg/s, 10 l/s or 10 Nl/s
Function description
USER INTERFACE → ADDITIONAL LINE → MULTIPLEX