Device Functions Proline Promass 83 MODBUS RS485 7 Block OUTPUTS
Endress+Hauser 71
VALUE F HIGH
MODBUS register:
Data type:
Access:
4205
3209
Float
read/write
! Note!
Function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function (4200).
Use this function to assign a variable to the END VALUE FREQUENCY(4203).
The value can be higher or lower than the value assigned to the VALUE F LOW.
Positive and negative values are permissible, depending on the measured variable
in question (e.g. mass flow). You define a measuring range by defining the VALUE
F LOW and VALUE F HIGH values.
User input: 5-digit floating-point number
Factory setting: Depends on nominal diameter [kg/h] or 2 [kg/l] or 200 [°C]
! Note!
Note that values with different signs cannot be entered for VALUE F LOW and
VALUE F HIGH, if SYMMETRY is the setting selected for the MEASURING
MODE function (4206). In this case, the message “INPUT RANGE EXCEEDED”
appears on the display.
A0004823
Fig. 26 : Behavior of frequency output
a = Measuring range
A = Frequency [%]
B = Measured variable (amount)
1 = Value F low
2 = Value F high
! Note!
Parameter setting examples for the frequency output → see overleaf.
Function description
OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (FREQUENCY)
125
100
a
B
A
0
1
2