7 Block OUTPUTS Device Functions Proline Promass 83 MODBUS RS485
70 Endress+Hauser
END VALUE
FREQUENCY
MODBUS register:
Data type:
Access:
4203
3205
Float
read/write
! Note!
Function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function (4200).
For defining a full scale frequency for the frequency output. You define the
associated measured value of the measuring range in the VALUE F HIGH function
(4205) described on Page 71.
User input: 5-digit fixed-point number: 2 to 10000 Hz
Factory setting: 10000 Hz
Example:
VALUE F HIGH = 10000 kg/h, full scale frequency = 10000 Hz:
i.e. a frequency of 10000 Hz is output at a flow of 10000 kg/h.
VALUE F HIGH = 3600 kg/h, full scale frequency = 10000 Hz:
i.e. a frequency of 10000 Hz is output at a flow of 3600 kg/h.
! Note!
In the FREQUENCY operating mode the output signal is symmetrical
(on/off ratio = 1:1). At low frequencies the pulse duration is limited to a
maximum of 2 seconds, i.e. the on/off ratio is no longer symmetrical.
VALUE F LOW
MODBUS register:
Data type:
Access:
4204
3207
Float
read/write
! Note!
Function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function (4200).
Use this function to assign a variable to the START VALUE FREQUENCY (4202).
The value can be higher or lower than the value assigned to the VALUE F HIGH.
Positive and negative values are permissible, depending on the measured variable
in question (e.g. mass flow). You define a measuring range by defining the VALUE
F LOW and VALUE F HIGH values.
User input: 5-digit floating-point number
Factory setting: 0 [kg/h] or 0 [kg/l] or 50 [°C]
! Note!
For a graphic illustration of the VALUE F LOW, see the VALUE F HIGH (4205)
function.
Note that values with different signs cannot be entered for VALUE F LOW and
VALUE F HIGH, if SYMMETRY is the setting selected for the MEASURING
MODE function (4206). In this case the message “INPUT RANGE
EXCEEDED” appears on the display.
The appropriate unit is taken from the following functions:
UNIT MASS FLOW (0400)
UNIT VOLUME FLOW (0402)
UNIT CORRECTED VOLUME FLOW (0404)
UNIT DENSITY (0420)
UNIT REFERENCE DENSITY (0421)
UNIT TEMPERATURE (0422)
(see Page 15 till Page 18).
Function description
OUTPUTS PULSE/FREQUENCY OUTPUT 1 CONFIGURATION (FREQUENCY)