Device Functions Proline Promass 83 MODBUS RS485 7 Block OUTPUTS
Endress+Hauser 69
Function description
OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → CONFIGURATION (FREQUENCY)
ASSIGN
FREQUENCY
MODBUS register:
Data type:
Access:
4201
3202
Integer
read/write
! Note!
Function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function (4200).
For assigning a measured variable to the frequency output.
Options:
0 = OFF
2 = MASS FLOW
5 = VOLUME FLOW
6 = CORRECTED VOLUME FLOW
7 = DENSITY
8 = REFERENCE DENSITY
9 = TEMPERATURE
Advanced selection: with the optional SW package CONCENTRATION
11 = TARGET MASS FLOW
12 = % TARGET MASS FLOW
13 = TARGET VOLUME FLOW
14 = % TARGET VOLUME FLOW
15 = CORRECTED TARGET VOLUME FLOW
16 = CARRIER MASS FLOW
17 = % CARRIER MASS FLOW
18 = CARRIER VOLUME FLOW
19 = % CARRIER VOLUME FLOW
20 = CORRECTED CARRIER VOLUME FLOW
21 = % BLACK LIQUOR
22 = ° BAUME
24 = ° API
25 = ° PLATO
26 = ° BALLING
27 = ° BRIX
28 = OTHER (_ _ _ _ flexible concentration)
Advanced selection: with the optional SW package ADVANCED DIAGNOSIS
51 = MASS FLOW DEVIATION
52 = DENSITY DEVIATION
53 = REFERENCE DENSITY DEVIATION
54 = TEMPERATURE DEVIATION
55 = TUBE DAMPING DEVIATION
56 = ELECTRODYNAMIC SENSOR DEVIATION
79 = OPERATING FREQUENCY FLUCTUATION DEVIATION
80 = TUBE DAMPING FLUCTUATION DEVIATION
Factory setting: MASS FLOW
! Note!
If you select OFF, the only function shown in the CONFIGURATION function
group is this function, in other words ASSIGN FREQUENCY (4201).
START VALUE
FREQUENCY
MODBUS register:
Data type:
Access:
4202
3203
Float
read/write
! Note!
Function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function (4200).
For defining an initial frequency for the frequency output. You define the
associated measured value of the measuring range in the VALUE F LOW function
(4204) described on Page 70.
User input: 5-digit fixed-point number: 0 to 10000 Hz
Factory setting: 0 Hz
Example:
• VALUE F LOW = 0 kg/h, initial frequency = 0 Hz:
i.e. a frequency of 0 Hz is output at a flow of 0 kg/h.
• VALUE F LOW = 1 kg/h, initial frequency = 10 Hz:
i.e. a frequency of 10 Hz is output at a flow of 1 kg/h.