Device Functions Proline Promass 83 MODBUS RS485 7 Block OUTPUTS
Endress+Hauser 91
Function description
OUTPUTS PULSE/FREQUENCY OUTPUT 1 OPERATION (STATUS)
ACTUAL STATUS
MODBUS register:
Data type:
Access:
4341
3248
Integer
read
! Note!
Function is not available unless the STATUS setting was selected in the
OPERATION MODE function (4200).
Use this function to check the current status of the status output.
Display:
0 = NOT CONDUCTIVE
1 = CONDUCTIVE
SIMULATION
SWITCH POINT
MODBUS register:
Data type:
Access:
4342
3249
Integer
read/write
! Note!
Function is not available unless the STATUS setting was selected in the
OPERATION MODE function (4200).
Use this function to activate simulation of the status output.
Options:
0 = OFF
1 = ON
Factory setting: OFF
! Note!
The “SIMULATION STATUS OUTPUT” message indicates that simulation is
active.
The measuring device continues to measure while simulation is in progress, i.e.
the current measured values are output correctly via the other outputs.
" Caution!
The setting is not saved in the event of a power failure.
VALUE
SIMULATION
SWITCH POINT
MODBUS register:
Data type:
Access:
4343
3250
Integer
read/write
! Note!
Function is not available unless STATUS was selected in the OPERATION MODE
function (4200) and the SIMULATION SWITCH POINT function (4342) is active
(= ON).
Use this function to define the switching response of the status output during the
simulation. This value is used to test downstream devices and the measuring
device itself.
Options:
0 = NOT CONDUCTIVE
1 = CONDUCTIVE
Factory setting: NOT CONDUCTIVE
" Caution!
The setting is not saved in the event of a power failure.