Device Functions Proline Promass 83 MODBUS RS485 7 Block OUTPUTS
Endress+Hauser 91
Function description
OUTPUTS → PULSE/FREQUENCY OUTPUT 1 → OPERATION (STATUS)
ACTUAL STATUS
MODBUS register:
Data type:
Access:
4341
3248
Integer
read
! Note!
Function is not available unless the STATUS setting was selected in the
OPERATION MODE function (4200).
Use this function to check the current status of the status output.
Display:
0 = NOT CONDUCTIVE
1 = CONDUCTIVE
SIMULATION
SWITCH POINT
MODBUS register:
Data type:
Access:
4342
3249
Integer
read/write
! Note!
Function is not available unless the STATUS setting was selected in the
OPERATION MODE function (4200).
Use this function to activate simulation of the status output.
Options:
0 = OFF
1 = ON
Factory setting: OFF
! Note!
• The “SIMULATION STATUS OUTPUT” message indicates that simulation is
active.
• The measuring device continues to measure while simulation is in progress, i.e.
the current measured values are output correctly via the other outputs.
" Caution!
The setting is not saved in the event of a power failure.
VALUE
SIMULATION
SWITCH POINT
MODBUS register:
Data type:
Access:
4343
3250
Integer
read/write
! Note!
Function is not available unless STATUS was selected in the OPERATION MODE
function (4200) and the SIMULATION SWITCH POINT function (4342) is active
(= ON).
Use this function to define the switching response of the status output during the
simulation. This value is used to test downstream devices and the measuring
device itself.
Options:
0 = NOT CONDUCTIVE
1 = CONDUCTIVE
Factory setting: NOT CONDUCTIVE
" Caution!
The setting is not saved in the event of a power failure.