The following screen is displayed:
1. Frame gauge reference
2. Calibration check
3. Calibration
Checking the calibration of the tracing table.
Before you start the calibration test, make sure that the reference of the frame gauge displayed on the
screen is the same as that marked on the gauge.
• If it is not identical, press to display the drop-down list: select the reference corresponding to the
gauge.
• If none of the references match that marked on the gauge, contact the technical department.
Select .
Insert the pattern gauge in the tracing table.
Press to start the test.
Wait: a timer is displayed for a few minutes.