78 WT.050.570.000.DE.IM.0714
Errors DEPOLOX
®
Pool
1.4
Error message Cause Remedy
Positioner Ym ? Ym range too narrow
Position 0 % or 100 % incorrectly ca-
librated
Set gap between 0 % and 100 % to
at least 50 % of the entire path
Positioner selected, but not connec-
ted
Check setting: Positioner with Ym
Feedback signal incorrect Check feedback (1 kOhm)
Positioner feedback incorrectly con-
nected or defective
Check ( refer to terminal diagrams)
Module ? Sensor module has been removed
Sensor module defective
Refit or replace the sensor module
Adaption ? Adaption terminated with error Refer to adaption
Overfeed ? Set maximum dosing time of a con-
trol output exceeded
Check cause, e. g. chemical tank
empty, check the dosing pump
DIN 19643 funct.? At least one of the values Cl
2
, pH,
Redox, Cl-N is outside the DIN spe-
cification
Check the control parameters
Run calibration
Flocculation ? pH value outside the flocculation li-
mit values
Check the flocculation limit value
Check pH measurement and control
Sample water insufficient (DI 1) Clean the filter
Feed delay active, e.g. during chan-
geover from manual to automatic
operation
Wait for feed delay to elapse or ter-
minate prematurely with the "Ack-
nowledge" key
high chlore
Rise time > max. delay or
µA-range of the sensor measuring mod-
ule exceeded during peak chlorination
Rise the max. delay time or select
higher measuring range.