13 Outputs
13.4 Activation of a binary output as pulse output FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
128
Enter the pulse width.
The range of possible pulse widths depends on the specification of the instrument (e.g.,
counter, PLC) that is to be connected to the output.
Press ENTER.
The max. flow that the pulse output can work with will be displayed now. This value is cal-
culated on the basis of the entered pulse value and pulse width.
If the flow exceeds this value, the pulse output does not work correctly. In this case, the
pulse value has to be increased.
Press ENTER.