17 Advanced functions
17.1 Totalizers FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
150
Automatic display toggling
The automatic toggling of the totalizer display between positive and negative flow direc-
tion can be set.
• Enter a time interval between 0 (off) and 5 s.
• Press ENTER.
Totalizer storing
It is possible to only store the value of the totalizer that is currently displayed or one value
for each flow direction.
• Select Special Funct.\SYSTEM settings\Storing\Quantity Storage.
• Press ENTER.
•If one is selected, only the value of the totalizer currently displayed will be stored. This
can apply for the positive and negative totalizer.
•If both is selected, the values of the totalizers for both flow directions will be stored.
• Press ENTER.
Totalizer behavior during the heat flow measurement
It is possible to output and store the values of the heat quantity totalizer and the volume
totalizer during heat flow measurement. Select
Special Funct.\SYSTEM settings\
Measuring\heat+flow quant.
• Select on to store and output the values of the heat quantity and volume totalizer during
the heat flow measurement.
• Press ENTER.
Special Funct.\SYSTEM settings\Measuring\Toggle totalizer
Special Funct.\SYSTEM settings\Storing\Quantity Storage
Special Funct.\SYSTEM settings\Measuring\heat+flow quant.