Multifunction Electrical Tester Calibrator
Supported SCPI Commands
137
[SOUR]:SAF:LOOP[:CURR]:PFC?
Description: This command returns the value of prospective fault current.
The Product returns the current value using exponential format.
Example: 100A is returned as 100e+00. This command switches
the Product into the Loop Impedance function mode.
Query: SAF:LOOP:PFC?
[SOUR]:SAF:LOOP:MAN(?) <DNPD>
Description: This command sets the manual entered value of residual
impedance.
Parameters: <DNPD> Parameter represents the value of residual
impedance in Ω. See the Specifications
(found online at www.Flukecal.com) for
acceptable ranges.
Example: SAF:LOOP:MAN 0.87 Set manual compensation value to 0.87 Ω.
Query: SAF:LOOP:MAN? Returns the manual compensation value.
[SOUR]:SAF:LOOP:SCAN?
Description: Returns the scanned loop residual impedance value. If not already
selected, it also switches the Product to the Loop Impedance
function.
Example: SAF:LOOP:MAN 0.87
Query: SAF:LOOP:SCAN? Product returns the manual entered value of
residual impedance using exponential
format. Example: 0.72 Ω is returned as
0.720e+00.
[SOUR]:SAF:LOOP:SCAN?
Description: Returns the scanned value of loop residual impedance.
Product returns the scanned value using exponential format.
Example: 0.72 Ω is returned as 0.720e+00. This command
switches the Product into the Loop Impedance function mode.
The Scan value is the same for Loop and Line functions.
Query: SAF:LOOP:SCAN?