[SOUR]:SAF:LOOP:PONS(?) <CPD> { ON | OFF }
Description: This command sets the Power ON (Start Up) switch. This
command switches the Product into the Loop Impedance function
mode. The state of the Power ON switch is the same for Loop and
Line functions.
Parameters: <CPD> Parameter represents the switch state.
ON Sets the power ON
OFF Sets the power OFF
Example: SAF:LOOP:PONS ON
Query: SAF:LOOP:PONS? The Product returns set switch state.
[SOUR]:SAF:LOOP:SRES(?) <DNPD>
Description: This command sets the series resistance of connection cables.
Parameters: <DNPD> Parameter sets the value of series resistance in Ω.
See the Specifications (found online at
www.Flukecal.com) for the acceptable ranges. This
command switches the Product into the Loop
Impedance function mode.
Example: SAF:LOOP:SRES 0.07
Query: SAF:LOOP:SRES? The Product returns the value of series
resistance using exponential format.
Example: 0.72Ω is returned as 0.720e+00.
[SOUR]:SAF:LOOP:PENR(?) <CPD> { ON | OFF }
Description: This command sets the PE-N resistance switch. This command
switches the Product into the Loop Impedance function mode.
Parameters: <CPD> Parameter represents the switch state.
ON Sets the PE-N resistance switch ON
OFF Sets the PE-N resistance switch OFF
Example: SAF:LOOP:PENR ON
Query: SAF:LOOP:PENR? The Product returns set switch state.
[SOUR]:SAF:LOOP:RESC
Description: Starts the internal measurement of power line residual impedance.
This command switches the Product into the Line Impedance
function mode.
Example: SAF:LOOP:RESC