Multifunction Electrical Tester Calibrator
Supported SCPI Commands
139
[SOUR]:SAF:VOLT[: LEV](?) [<DNPD>]
Description: This command sets the output voltage from the Voltage calibrator
function.
Parameters: <DNPD> Parameter represents the value of
generated voltage expressed in Volts.
Example: SAF:VOLT 100 Voltage Calibrator 100V.
Query: SAF:VOLT? The Product returns the value of the
generated voltage using exponential format.
Example: 50.5 V is returned as 50.50e+00.
[SOUR]:SAFE:VOLT:CURR?
Description: This command returns the measured load current in the voltage
calibrator function. The Product returns the current value using
exponential format. Example: 5 mA is returned as 5e-03. This
command switches the Product into the Voltage Calibrator function
mode.
Query: SAF:VOLT:CURR?
[SOUR]:SAF:VOLT:FREQ(?) <DNPD>
Description: This command sets the frequency in the AC Voltage Calibrator
mode. This command switches the Product into the Voltage
Calibrator function mode.
Parameters: <DNPD> Parameter represents the value of
frequency expressed in Hz.
Example: SAF:VOLT:FREQ 60 60 Hz.
Query: SAF:VOLT? The Product returns the value of frequency
using exponential format. Example: 50 Hz is
returned as 50.0e+00.
[SOUR]:SAF:VOLT:FUNC(?) <CPD> {DC | AC | SYNC }
Description: This command selects dc or ac output voltage signal from the
Voltage calibrator function. This command switches the Product
into the Voltage Calibrator function mode.
Parameters: <CPD> DC sets a dc output signal
AC sets an ac output signal
SYNC sets an ac output signal synchronized with
the power supply frequency
Example: SAF:VOLT:FUNC AC Set to an ac voltage output signal.
Query: SAF:VOLT:FUNC? The Product returns AC, DC, or SYNC.