Multifunction Electrical Tester Calibrator
Supported SCPI Commands
141
[SOUR]:SAF:MET:POW?
Description: This command returns the measured power. The Product returns
the power value using exponential format. Example: 45 VA is
returned as 45.00e+00. This command switches the Product into
the Meter function mode.
Query: SAF:MET:POW?
[SOUR]:SAF:MET:UNIT <CPD> { VA | VAR | W }
Description: This command sets the units when measuring power. If not
already selected, this command switches the Product into the
Meter Multimeter mode.
Parameters: <CPD> VA sets the units to VA (apparent power)
VAR sets the units to VAR (reactive power)
W sets the units to W (active power)
Query: SAF:MET:UNIT?
[SOUR]:SAF:MET:VOLT?
Description: This command returns the measured voltage. The Product returns
the voltage value using exponential format. Example: 230 V is
returned as 230.0e+00. This command switches the Product into
the Meter function mode.
Query: SAF:MET:VOLT?
[SOUR]:SAF:MET:PHAS?
Description: This command returns the measured phase.
The Product returns the phase value using exponential format.
Example: 60° is returned as 60.00e+00. This command switches
the Product into the Meter function AC mode.
Query: SAF:MET:PHAS?
[SOUR]:SAF:HIPL:FUNC(?) <CPD> {DC | AC}
Description: This command selects dc or ac hipot leakage current mode. This
command switches the Product into the Hipot leakage current
function mode.
Parameters: <CPD> DC sets a dc hipot leakage current mode.
AC sets an ac hipot leakage current mode.
Example: SAF:HIPL:FUNC AC Set to the ac hipot leakage current mode.
Query: SAF:HIPL:FUNC? The Product returns ac or dc.