EasyManuals Logo

Fluke NetDAQ 2640A User Manual

Fluke NetDAQ 2640A
353 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #179 background imageLoading...
Page #179 background image
Maintenance
Performance Test
6
6-7
Table 6-2. Recommended Test Equipment
Instrument Type Minimum Specifications Recommended Model
Multifunction DC Voltage: Fluke 5700A
Calibrator
Range: 90 mV to 300V dc
Accuracy: 0.002%
AC Voltage:
Frequency Voltage Accuracy
1 kHz 29 mV to 300V 0.05%
100 kHz 15 mV to 300V 0.5%
Frequency:
10 kHz 1V rms 0.01%
Ohms:
Ohms Accuracy
190Ω 0.005%
1.9 kΩ 0.005%
19 kΩ 0.005%
190 kΩ 0.005%
1.9 MΩ 0.005%
Mercury Thermometer 0.02°C Resolution Princo ASTM-56C
Thermocouple Probe Type T Fluke P20T
Oil/Water Bath Thermos bottle and cap
Digital Multimeter General Purpose Measurement Fluke 77
Signal Generator Sine wave. 0.5 to 1V rms, 10 Hz to 5 kHz Philips PM5193
Alternative Equipment List
Instrument Type Recommended Model
DC Voltage Calibrator Fluke 5440B
DMM Calibrator Fluke 5100B (AC Volts Only)
Function/Signal Generator Philips PM5139 or HP33120A
Decade Resistance Source General Resistance RDS 66A