EasyManuals Logo

Fluke NetDAQ 2640A User Manual

Fluke NetDAQ 2640A
353 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #249 background imageLoading...
Page #249 background image
RTD Linearization
Custom-385 RTD
D
D-5
2645A
2640A
-100 0 100 200 300 400 500 600
Temperature - Degrees C
Accuracy Envelope: Custom RTD, 392 Probe
1 Year Slow: 18-28 Degrees C
Alpha: 0.0039261 Delta: 1.49512239 Beta: 0.10188228
-1.5
-0.5
0.0
0.5
1.0
Error - Degrees C
1.5
-1.0
Figure D-4. 392 RTD