2640A/2645A NetDAQ
Users Manual
J-4
A/D Reference Balance Values Invalid! Continue?
Your instrument has detected a fault during internal performance monitoring.
This message does not appear under normal operating conditions. Note the
message and click Yes to continue operations or No to terminate operations. If the
fault recurs, stop operations, and cycle instrument power. If the fault still persists,
remove the instrument from service and return it to a Fluke Service Center for
repair, along with a description of the error message.
A/D Zero Offsets Invalid! Continue?
Your instrument has detected a fault during internal performance monitoring.
This message does not appear under normal operating conditions. Note the
message and click Yes to continue operations or No to terminate operations. If the
fault recurs, stop operations, and cycle instrument power. If the fault still persists,
remove the instrument from service and return it to a Fluke Service Center for
repair, along with a description of the error message.
All associated configurations will be lost. Continue?
You are deleting all the instruments in the Communications Configuration File,
which will also clear the configuration data from the instruments and remove all
the instruments from the Icon Bar and network.
Clearing the Communications Configuration File of all network instruments
invalidates all references to network instruments in all setup files as the
Communications Configuration File is common to all setup files. Click Yes to
proceed or No to cancel the operation.