EasyManua.ls Logo

Frick QUANTUM LX - Current Zone Configuration; Calibrating Sensors; Establishing Setpoints; Cooling Setpoints

Frick QUANTUM LX
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
QUANTUM™ LX EVAPORATOR CONTROL PANEL
OPERATION
090.610-O (MAR 2016)
Page 19
Current Zone Conguration
Access this screen from the Main Menu. Select Conguration,
then Current Zone Conguration. The following screen will
appear:
Notice in the upper right hand corner of this screen is a set-
point box that contains a value. You may change the value in
this box to the number of the zone that you wish to congure.
You must then press the [Submit] button for the screen to dis-
play that zone. In the above example, the number 1 is shown
on the box. Notice that right below the setpoint box, the words
ZONE #1 is displayed, so Zone 1 is the zone parameters that
are being shown.
Many other screens also show the Zone selection box in the
upper right corner. Any screen that displays this box presents
data that pertains exclusively to the selected zone. The selec-
tions on the Current Zone Conguration page must be made
for each Evaporator zone that has been enabled on the Global
Zone Conguration page.
The Current Zone Conguration page permits the user to en-
able any four control functions:
Cooling
• Defrost
Dehumidication
Heating
The Options section allows the user to congure various me-
chanical aspects of the system, which include the type and
operation of valves as well as the fan type.
CALIBRATING SENSORS
Each temperature or humidity sensor in the Evaporator should
be calibrated before the system is put into operation.
Access this screen from the Main Menu. Select Calibration,
then Analog Inputs:
The process used for establishing calibration values is rela-
tively simple. It will however require a container of ice water
or an infrared digital temperature probe.
First, establish the temperature for a specic zone’s tempera-
ture sensor. If a digital infrared measurement device is used,
simply enter the calibrated temperature into the Current Value
box and press [Submit]. The system will automatically cal-
culate the correct offset value to use. If an ice bath is used,
simply allow the temperature of the sensor to stabilize, and
then enter 32 into the Current Value box and press [Submit].
This process should be repeated for each sensor in the system.
ESTABLISHING SETPOINTS
Cooling Setpoints
Access this screen from the Main Menu. Select Set-
points, then Cooling, and nally Mode (1-4):

Table of Contents

Other manuals for Frick QUANTUM LX

Related product manuals