AT600 User’s Manual 99
Chapter 5. Communication
Command 160 (0xA0): Read Analog Measurement Range Values (cont.)
Command 161 (0xA1): Read Loop Current Error Handling
This command is to read the loop current output error handling.
Table 44: Command-Specific Response Codes for Read Analog Measurement Range Value
Code Class Description
0 Success No Command-Specific Errors
1 - 5 Undefined
6 Error Device-Specific Command Error
7 - 127 Undefined
Table 45: Request Data Bytes for Read Loop Current Error Handling
Byte Format Description
None
Table 46: Response Data Bytes for Read Loop Current Error Handling
Byte Format Description
0 Unsigned-8 Analog Output Error Handling:
0: Low;
1: High;
2: Hold;
3: Other value;
1 - 4 Float Error Value, unit is mA
Table 47: Command-Specific Response Codes for Read Loop Current Error Handling
Code Class Description
0 Success No Command-Specific Errors
1 - 5 Undefined
6 Error Device-Specific Command Error
7 - 127 Undefined