AT600 User’s Manual 57
Chapter 3. Initial Setup and Programming
3.7.1 Programming the Meter Setup
3.7.1a Setting Zero Cutoff
Near a zero flow rate, the AT600’s readings may fluctuate due to small offsets caused by thermal drift or similar
factors. To force a Zero display reading, when there is minimal flow, enter a zero cutoff value as described below.
Select
Meter Setup and press []. The following screen appears.
Select
Zero Cutoff and press []. The following screen appears.
Press
[] again, and a screen similar to the following appears.
Note: The units that appear will be the units selected in Units Setting on page
28.
To change the
Zero Cutoff, use the [] or [] arrow key to select the digit to be
changed, use the [] or [] arrow keys to provide a new number and press
[]
to save. Press [] to return to the previous screen.
Sensor Setup
Meter Setup
Pipe
Transducer
Traverses
Meter Setup
Zero Cutoff
Meter Setup/Zero Cutoff
Zero Cutoff
0.05 m/s
Zero Cutoff
UNIT: m/s
0.05
]MOVE
]UNDO [ ]SAVE
]
]
]MODF
[