Appendix A. Specifications
156 AT600 User’s Manual
A.3 Electronics
A.3.1 Enclosures
Epoxy coated, copper free, aluminum
A.3.2 Weatherproof
Enclosures: IP67
Please contact sales representative for other transducers.
A.3.3 Electronics Classifications (Pending)
CE (EMC directive) IEC 61326-1:2012, IEC 61326-2-3:2012, LVD 2006/95/EC, EN 61010-1 2010)
ETL(UL61010-1, CSA 22.2 No 61010.1, No. 142, FCC part 15, CISPR 11)
WEEE Compliance
ROHS Compliance
Note: The electronics package includes an installed battery which shall only be replaced at a GE Service center.
Replacement involves de-soldering battery contacts, which could lead to a breach of Functional Safety. Please
contact GE Service to get this battery replaced.
A.3.4 Display Languages
English/Chinese/German/French/Italian/Japanese/Portuguese/Russian/Spanish
The meter will be set into the language requested by customer before shipping to customer.
A.3.5 Keypad
Film keypad, Six-button keypad, for full functionality operation
A.3.6 Inputs/Outputs
Standard: One analog output*, service (RS485) output, two digital outputs***, one gate input;
Option A: One analog output* with HART**, service (RS485) output, two digital outputs***, one gate input;
Option B: One analog output*, service (RS485) output, one Modbus(RS485) output, two digital outputs***, one gate
input;
*Analog output is NAMUR NE43 compliant
**HART is compliant with Protocol of Version 7
***Digital Outputs are programmable as either pulse, frequency, alarm, or control outputs. Digital outputs will be
configured into the output mode as requested by customer before shipping to customer.