Chapter 5. Communication
138 AT600 User’s Manual
Command 225 (0xE1): Read Signal Setup (cont.)
Command 226 (0xE2): Read Flowmeter S/N
This command is to read flow meter s/n.
Table 158: Command-Specific Response Codes for Read Signal Setup
Code Class Description
0 Success No Command-Specific Errors
1 Undefined
2 Error Invalid Selection
3-4 Undefined
5 Error Too Few Data Bytes Received
6 Error Device-Specific Command Error
7-127 Undefined
Table 159: Request Data Bytes for Read Flowmeter S/N
Byte Format Description
0 Unsigned-8 Flowmeter S/N:
1. Electronic S/N
2. UP Sensor
3. S/N
4. DN Sensor S/N
Table 160: Response Data Bytes for Read Flowmeter S/N
Byte Format Description
0 Unsigned-8 signal setup type:
1. Electronic S/N
2. UP Sensor
3. S/N
4. DN Sensor S/N
1 - 16 Unsigned-8 S/N