Chapter A4: Image Acquisition
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) A4-11
Operator Manual © 2008 General Electric Company. All rights reserved.
Figure A4-7 Cassette exam Acquisition screen
Follow this process to conduct a Cassette Exam.
NOTE: Make sure that the GE Portable DR Detector is in the detector bin before beginning a cassette
exam. Exposures will be inhibited if the detector is placed in the bin after the exam has begun. To
clear the inhibit, either:
– Select a different view and then reselect the view you wish to perform as shown in step 4 be-
low (recommended).
– Remove the detector from the bin for the duration of the exam.
Refer to Chapter 3: Hardware Overview-Unit Features
(p. 3-1) for more information about the
detector bin.
1. Press [CASSETTE EXAM] from the bottom of the Worklist screen.
NOTE: Because a cassette exam does not use digital image storage, you do not select procedures from
the Worklist.
Figure A4-8 [CASSETTE EXAM] button on Worklist
The Select Protocols screen appears.
2. Select the protocols to perform.