Chapter 4: Moving and Positioning
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) 4-6
Operator Manual © 2008 General Electric Company. All rights reserved.
Figure 4-7 Lock release mechanism
Adjust Column and Telescoping Arm
In the following steps you can free the Definium AMX 700 components for positioning in one of two ways:
by gripping the collimator handle(s) to keep the lock system released or, whenever necessary, overriding
the lock system with slight force.
When possible, we recommend that you use the lock release system.
IMPORTANT!: Remove the tether from the holder before rotating the column.
1. Rotate the arm and vertical column assembly to horizontally position the telescoping arm. The
vertical column can be rotated 270 degrees clockwise or counterclockwise (detent at 0 degree park
location). See Figure 4-8.
2. Raise the arm on the vertical column to the desired height.
3. Extend the retractable arm to the desired length and release. See Figure 4-8.
Figure 4-8 Extending the retractable arm
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).