Chapter 7: Maintenance and Service
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) 7-1
Operator Manual © 2008 General Electric Company. All rights reserved.
Chapter 7: Maintenance and Service
Topics covered include:
• Cleaning the Unit
(p. 7-1)
• Periodic Maintenance by Service Personnel
(p. 7-1)
– Qualified Service Available
(p. 7-2)
– Maintenance Schedule
(p. 7-2)
Cleaning the Unit
We recommend that you clean the Definium AMX 700 regularly with a soft cloth and a mild, non-abrasive
cleaner. Apply the cleaner to the cloth, not directly to the unit, to safe-guard against damage to the
circuitry. If a corrosive chemical is spilled or splashed on the unit surface, be sure to clean it off
immediately.
The following chemicals and products have been tested and approved by GE for cleaning the Definium
AMX 700 and GE Portable DR Detector.
•Cidex OPA
•Gigasept FF
•T-Spray
•Inciden Plus
• Perasafe
• Ster-Bac Blue
• Super Sani-Cloth Germicidal Wipes
• TriGene
• Isopropyl Alcohol (70% maximum solution)
•Bleach (50% solution)
Refer to Chapter A1: Detector Overview-Cleaning
(p. A1-10) for more information about cleaning the GE
Portable DR Detector and its accessories.
Periodic Maintenance by Service Personnel
To insure the continued safe performance of your Definium AMX 700 portable x-ray unit, establish a
periodic maintenance program with a qualified service representative.
Periodic maintenance checks are required thirteen months after installation and every year thereafter. A
periodic maintenance schedule and a sample mAs or kVp accuracy chart are provided on the following
pages.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).