Chapter 6: Charging the Batteries
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) 6-3
Operator Manual © 2008 General Electric Company. All rights reserved.
Recharge Recommended
When battery charge dips below approximately 10%, recharge is recommended.
As the Definium AMX 700 is used in its various modes of operation, the number of segments lit will slowly
decrease.
It’s important to note that you do not have to raise your kVp and mAs value to compensate for low
battery capacity.
To ensure maximum battery life, the Definium AMX 700 should not be recharged until at least one-half of
the bar graph segments have extinguished.
Battery Drain
Figure 6-2 shows how much of the battery capacity is used for the various Definium AMX 700 operating
modes in a relative basis.
Figure 6-2 Battery drain
As Figure 6-2 demonstrates, the majority of the battery pack’s capacity is used for drive and idle. Thus,
reducing the amount of drive time by parking the Definium AMX 700 in a central location and turning the
system OFF when not in use will make more capacity available for x-ray generation and will increase the
service life of the batteries.
Bar Graph Display vs. Battery Loading
Table 6-2 details the various battery loads and how they affect the bar graph display.
Table 6-2 Battery loads
Load mode Approximate discharge rate Typical daily usage*
Drive (Drive handle engaged) 0.45 segments/minute 30 segments
17%
71%
8%
2%
2%
Idle Field
Light
X-ray
Generation
X-ray
Prep
Drive
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).