Chapter A5: Image Viewer
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) A5-17
Operator Manual © 2008 General Electric Company. All rights reserved.
In the event of incorrect automatic shuttering, the image can be recovered by manually re-shuttering the
image to visualize the desired anatomy and then re-processing with the same look.
Re-process Images
Image re-processing allows the system to extract more information from an already acquired image by
changing the processing settings instead of taking additional exposures.
Re-processing can be performed on any image that has a corresponding raw data set. Images can be re-
processed both in live exams and in review mode.
Re-processing creates a new image in the “PROCESSED” series.
NOTE: When closing an exam or closing patient in review mode, you must select to save changes to
images or the re-processed images will not remain in the series. Refer to Save Changes to Images
(p. A5-27) for more information.
The initial image processing is determined by the default that is configured for the protocol. Refer to
Chapter A7: Set Preferences-Image Processing
(p. A7-31) for more information.
Table A5-6 describes the settings used to re-process an image.
Figure A5-13 Reprocessing panel
Table A5-6 Reprocessing tool descriptions
Function Description
Anatomy Changes the anatomical region.
View Changes the view.
Image Type Displays the image type of the acquisition.
Images cannot be re-processed with a different image type than originally
acquired.
Patient size Changes the patient size.
FOR TRAINING PURPOSES ONLY!