Chapter A1: Detector Overview
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) A1-2
Operator Manual © 2008 General Electric Company. All rights reserved.
Detector Functions
Detector primary functions are:
• To convert x-ray data into digital image data
• To transfer the digital data to an external workstation for processing and display
Description
The detector is an x-ray imaging device. It consists of an array of 2022 x 2022 pixels (40.4 x 40.4 cm).
Each pixel is attached to a data acquisition circuit that converts incoming x-ray signal to 14-bit digital
data.
The detector is constructed from a metal case. The front face contains a graphite x-ray imaging window.
The back face contains the usage instructions and safety warnings.
Figure A1-1 shows the front of the detector, which faces toward the x-ray source.
Figure A1-1 Detector front
1. LED indicator lights
2. Handle
3. Tether
4. Detection area (inside white marker lines)
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).