Chapter A7: Set Preferences
Definium AMX 700 X-Ray System 5161515-1EN Rev. 6 (10 February 2008) A7-34
Operator Manual © 2008 General Electric Company. All rights reserved.
Table A7-10 Image Processing parameters
Parameter Definition
Contrast
Adjust
Adjusts image contrast. Also known as window width.
Move the slider right for more contrast (towards pure black and white).
Move the slider left for less contrast (towards uniform gray).
Range: 25% – 400% with increment of 1%
Brightness
Adjust
Adjusts image brightness from 25-210%. Also known as window level.
Move the slider right for a lighter image.
Move the slider left for a darker image.
NOTE: Brightness Adjust and Contrast Adjust should not be used for per-image
tuning/correction of brightness and contrast. They are used to modify the
default look of images from a particular exam. For example, if images from a
particular type of exam are appearing consistently lighter than they should
be, the Brightness Adjust slider can be moved left to adjust. However, for small
alterations of brightness on a particular image, use the Brightness and
Contrast controls in the Image Viewer screen.
Tissue
Contrast
Controls the general contrast between the thick and thin anatomy.
Move the slider right to decrease bone/soft-tissue contrast.
Move the slider left to increase bone/soft-tissue contrast.
Range: –0.15 to 0.15 with 0.01 increments.
NOTE: Unlike Tissue Equalization, which controls contrast within under-penetrated
(thick) or over-penetrated (thin) areas, Tissue Contrast controls the general
contrast between the thick and thin anatomy. For example, Tissue Contrast
can be used in combination with Tissue Equalization to define a Chest PA look
that is equalized in the lung fields, but with light spine/ribs.
Edge The amount of detail visible in bone structures. Increased edge equals increased
detail.
Move the slider right to make images sharper.
Move the slider left to make images smoother.
Range: 1 to 10 (discrete setting) with increments of 1